Open original FDA record ↗Recall record
- FDA identifier
- Z-1019-2010
- Company in FDA record
- Siemens Healthcare Diagnostics Inc
- FDA classification date
- Mar 8, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Software: Reference sensor failure may result in inaccurate pH reuslts
Source facts
- Action
- An "Urgent Field Safety Notice" dated September 2009 was sent via overnight mail to customers and included software Version 3.5.2 with instructions for install that will detect and prevent the reporting of potentially errant test results. The customers are requested to please install the included software update immediately and return the attached Fax back form. Siemens will provide the mandatory update to all customers with the Rapidpoint 400 /405 test system. If you have any questions, please contact your local Technical Support at 508-359-3877.
- Initiated
- 2009-10-30
- Root cause
- Component design/selection
- Terminated
- 2015-09-04
- Distribution
- Worldwide distribution: USA and Australia, Austria, Bahamas, Barbados, Belgium, Bosnia Herzeg, Brazil, Canada, Canary Islands, Chile, Croatia, Czech Republic, Egypt, France, Germany, Holy See (Vatican City State) Hong Kong, Ireland, Italy, Japan, Macedonia, Malaysia, M¿xico, New Zealand Not assigned P.R. China, Poland, Portugal, Puerto Rico, Refurb.Systems U.S.A Republic Korea, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, U.A.E., United Kingdom
- Recall number
- Z-1019-2010
- Recall status
- Terminated
- CDRH record ID
- 85805
- FDA classified
- 2010-03-08
- Recall event number
- 53541