- Action
- On 1/30/2009, Philips began sending out URGENT-Device Correction PCR Eleva notifications to their consignees.
The consignees were informed of the hazards as when a patient is selected in patient list and examination tab pressed with background query running at the same time, there would be a risk that the wrong patient is selected.
The firm issued the Field Corrective Order (FCO) 73200029 to their representative to correct the problem. The corrective action would be a software upgrade that resolve the issue.
All consignees can contact Philips Call Center at 800-722-9377, #5, #2 and reference "FCO 73200029" with any questions.
- Initiated
- 2008-12-23
- Root cause
- Software design
- Terminated
- 2011-08-25
- Distribution
- Nationwide Distribution -- USA, including states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WI, WV, and WY and to LANDSTUHL ARMY BASE, APO in United Arab Emirates.
- Recall number
- Z-1019-2011
- Recall status
- Terminated
- CDRH record ID
- 96561
- FDA classified
- 2011-02-04
- Recall event number
- 57558