Skip to content

Public FDA record

Z-1019-2017: OPT Croutons is a synthetic bone substitute designed for the repair of bony defects and is made of pure b

Recall record Z-1019-2017 for Acumed LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1019-2017
Company in FDA record
Acumed LLC
FDA record creation date
Aug 23, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

OPT Croutons, part number 65-0210-S, is recalled due to the result of inadequate documentation of the sterilization dose used in production. Product sterilized at the site did not receive the minimum dose required per the new performance qualification.

Source facts

Action
Acumed sent the Urgent Recall Notice letter, dated December 1, 2016, to US customers on December 5, 2016. Customers are advised to follow the instructions provided by Acumed: IMMEDIATELY stop using and/or distributing the product and complete the following actions: 1. Perform a physical count of your inventory. 2. Remove the affected OPT Croutons, such that the croutons are out of service. 3. Record this data on the survey found following this recall letter. 4. Fill out the survey and return it via email to RecaiiNotification@acumed.net. 5. For any product that needs to be returned: Domestic Customers: please contact Acumed Agent Inventory via email at Agentlnventory@acumed.net or contact your agent directly to be assigned a Return Material Authorization (RMA) number and applicable shipping information. International Customers: please contact Acumed Business Services via email at BusinessServices@acumed.net or via phone at (877) 627-9957 Option 4 (US and Canada) or +001 (503) 627-9957 and choose Option 4 (outside of US and Canada) to be assigned a Return Material Authorization (RMA) number and applicable shipping information. 6. Once completed, please return the OPT Croutons that have been removed to Acumed using the Return Label provided by Acumed and included in your replacement shipment. Please contact Acumed Business Services via email at BusinessServices@acumed.net or via phone at (877) 627-9957 Option 4 (US and Canada) or +001 (503) 627-9957 and choose Option 4 (outside of US and Canada) with any questions regarding this notification.
Initiated
2016-12-05
Root cause
Vendor change control
Terminated
2017-03-22
Distribution
distributed in FL, IN, KS, MI, MO, NE, OH, PA, TX and Peurto Rico.
Recall number
Z-1019-2017
Recall status
Terminated
CDRH record ID
151650
FDA record created
2017-08-23
Recall event number
75904

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗