Open original FDA record ↗Recall record
- FDA identifier
- Z-1019-2023
- Company in FDA record
- SunMed Holdings, LLC
- FDA classification date
- Jan 26, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is an orogastric (OG) tube size discrepancy between the labeling in the IFU and the printed text on the device. If the clinician is unable to pass a gastric catheter through one of the channels, the clinician may use a smaller-diameter gastric catheter.
Source facts
- Action
- On December 22, 2022, the firm notified affected customers through Urgent Medical Device Product Correction letters. Customers were informed that there is a discrepancy in the orogastric (OG) tube size indicated in the labeling in the instructions for use (IFU) versus the printed text on the device. Customers were provided with the correct recommended maximum OG tube/catheter sizes for each impacted Air-Q3G and Air-Q SPG model. If product has been further distributed, customers should forward the recall notice. If customers have questions about this recall notification, please contact Rob Yamashita, VP of Regulatory Affairs, at 616-259-8400, toll free at 800-433-2729, or ryamashita@sun-med.com.
- Initiated
- 2022-12-22
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution in the states of AL, CA, FL, MI, OH, TN, TX, WA.
- Recall number
- Z-1019-2023
- Recall status
- Open, Classified
- CDRH record ID
- 197607
- FDA classified
- 2023-01-26
- Recall event number
- 91391