Z-1019-2025: Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatien

Recall record Z-1019-2025 for Fresenius Kabi USA, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1019-2025
Company in FDA record
Fresenius Kabi USA, LLC
FDA classification date
Feb 4, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.

Source facts

Action
Fresenius Kabi began notifying consignees on 01/10/2024 via emailed letter. Consignees were instructed to check if your device contains the affected software, review and become familiar with the temporary Risk Mitigation actions for each anomaly outlined in the letter and notify customers/personnel if the product was transferred or further distributed. Additionally, to facilitate the installation of the pump software, LVP SW 5.10, your facility must install the new Ivenix Infusion Management System (IMS) software version (5.2) on your IMS server and install the new software version (5.10) for the Ivenix Large Volume Pump (LVP). Please post this letter in areas that affected devices are stored and used such that users of this device are aware of the issues and recommended risk mitigation measures. Lastly, consignees were requested to complete and return the Customer Reply Form.
Initiated
2025-01-10
Root cause
Software Design Change
Distribution
US distribution to CA, CO, GA, ID, MI, MS, NE, NJ, NV, OK, TX, UT, VA, WA, WI.
Recall number
Z-1019-2025
Recall status
Open, Classified
CDRH record ID
211967
FDA classified
2025-02-04
Recall event number
96085

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗