Z-1020-2025: HemosIL Heparin Calibrators; Part Number: 0020300600; in vitro diagnostic

Recall record Z-1020-2025 for Instrumentation Laboratory. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1020-2025
Company in FDA record
Instrumentation Laboratory
FDA classification date
Jan 24, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Control indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range.

Source facts

Action
On December 18, 2024 URGENT MEDICAL DEVICE CORRECTION letters were email to customers. Mandatory Customer Actions Based on the above, please take the following immediate actions: - Check inventory for lots of HemosIL Heparin Calibrators (Part No. 0020300600). See Appendix A for a list of in-date lots with corrected value assignments for the HEP CAL 2 and HEP CAL 3 levels. - Follow the instructions below to update the ACL TOP test parameters (same instructions for all models): - Select Setup from the menu bar and then select Test List from the dropdown menu. - From Test List , double click on the Test Code Anti-Xa to open Test Definition Anti-Xa . - From Test Definition Anti-Xa, click on Pre-diluted calibrators under the header Calibration . An editable table titled Calibration points will appear with HEP CAL 1, HEP CAL 2, and HEP CAL 3 and their predefined Target Values IU/mL of 0.00, 0.80 and 2.00, respectively. - In the Calibration points table, change the target values for HEP CAL 2 and HEP CAL 3 to the lot-specific calibrator values provided in Appendix A. - After entering the corrected HEP CAL 2 and HEP CAL 3 values, click on the save icon (picture of floppy disk) to save the changes. - Recalibrate using the HemosIL Liquid Anti-Xa assay lot in use after making the test parameter update. - Run quality controls per labeled instructions for use. If needed, determine control ranges for HemosIL UF Heparin Controls (Part No. 0020300300) and HemosIL Low Molecular Weight Heparin Controls (Part No. 0020300200) in accordance with your internal protocol until the new acceptance ranges are available. - Repeat the above test parameter update, recalibration, and quality control for all ACL TOP instruments. - Share this information with your laboratory staff. - Forward this notification to all affected locations within your facility. - Retain a copy of this notification for your records. - Complete and sign the enclosed Mandatory Response Tracking Form a
Initiated
2024-12-18
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide and the countries of Andorra, Algeria, Armenia, Australia, Austria, Belarus, Belgium, Brazil, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Cyprus, Czechia, Denmark, Finland, France, Georgia, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Iran (Islamic Republic of), Ireland, Israel, Italy, Japan, Korea (the Republic of), Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Macao, Malaysia, Mexico, Netherlands (the), Norway, Philippines (the), Poland, Portugal, Qatar, Romania, Russian Federation (the), Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan (Province of China), Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Viet Nam.
Recall number
Z-1020-2025
Recall status
Open, Classified
CDRH record ID
211851
FDA classified
2025-01-24
Recall event number
96044

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗