Skip to content

Public FDA record

Z-1021-2025: Design Options COMBINED SPINAL/EPIDURAL TRAY- Intended for administration of analgesic/anesthetic agent

Recall record Z-1021-2025 for B. Braun Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1021-2025
Company in FDA record
B. Braun Medical, Inc.
FDA classification date
Jan 24, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Incorrect catheter connector within the kit resulting in inability to provide a secure connection to the epidural catheter. The incorrect catheter connector is smaller than the required catheter connector and therefore is unable to be used to administer medication.

Source facts

Action
B Braun issued Urgent Medical Device Correction Notification letter on 12/04/24 via USPS Certified Mail. Letter states reason for recall, health risk and action to take: 1. Review the Medical Device Correction Notification in its entirety and ensure that all users in your organization of the above-mentioned product, and other concerned persons are informed about this voluntary correction. If you have further distributed the affected product, please forward this notice to your consignees. The correction is to be extended to the hospital/healthcare facility level. 2. Post this notification where the affected products are stored. Additional catheter connectors are being supplied to you and should be stored where the affected inventory is utilized. In the event that an incorrect catheter connector is identified, please utilize these additional devices to facilitate replacement. 3. Returned the completed Urgent Medical Device Correction Acknowledgement Form to B. Braun Medical Inc. Quality Assurance department by faxing the form to (610) 849-1197 or email to recalls@bbraunusa.com within two (2) weeks of receipt, even if the total inventory in your possession is zero (0). Adverse reactions or quality problems experienced with this product, or questions about this recall may be reported to BBMI s Postmarket Surveillance Department by calling 1-844-903-6417.
Initiated
2024-12-04
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the states of MO, OK.
Recall number
Z-1021-2025
Recall status
Open, Classified
CDRH record ID
211427
FDA classified
2025-01-24
Recall event number
95874

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗