Open original FDA record ↗Recall record
- FDA identifier
- Z-1022-2013
- Company in FDA record
- Genetic Testing Institute,inc
- FDA classification date
- Mar 28, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Lifecodes PF4 Enhanced assay, lot number 3000389, is being recalled because of observed inconsistent replicate values when testing the kit Positive Control and patient samples. The variation in reactivity can lead to invalid assays. If an invalid assay is obtained, those results should not be reported and the test should be re-run. An invalid assay would not result in significant harm to a pati
Source facts
- Action
- Hologic/Gen-Probe sent an Urgent Medical Device Recall letter dated February 4, 2013, via FedEx overnight deliver yto all affected customers. The letter identified affected product, stated the issue, and asked that product from the affected lot stop being used. Reported results for all samples tested using the affected lot should be reviewed. The referring physician should be notified that any negative patient result may be erroneous. Remaining inventory should be returned to Gen-Probe. A response form was also asked to be returned. For questions customers were instructed to contact Technical Support at 262-754-1000 or waukeshatechsupport@gen-probe.com. For questions regarding this recall call 262-754-1026.
- Initiated
- 2013-02-04
- Root cause
- Process control
- Terminated
- 2016-03-21
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Canada, China, Slovakia, and Spain.
- Recall number
- Z-1022-2013
- Recall status
- Terminated
- CDRH record ID
- 116359
- FDA classified
- 2013-03-28
- Recall event number
- 64501