Z-1022-2025: Medtronic PIN, 100mm STERILE PERCUTANEOUS REF: 9733235 Medtronic PIN, 150MM, STERILE PERCUTANEOUS REF: 9

Recall record Z-1022-2025 for Medtronic Navigation, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1022-2025
Company in FDA record
Medtronic Navigation, Inc.
FDA classification date
Jan 27, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to an increase in complaint that the percutaneous reference pin would not fit into the patient reference frame or percutaneous pin adapter.

Source facts

Action
On 12/04/2024, the firm send an "URGENT: MEDICAL DEVICE RECALL" Letter via UPS 2-Day mail to customers to inform them that Medtronic has become aware that certain percutaneous pin lots have been identified as having the potential for an out of round diameter, that may render the percutaneous pin unable to fit into the Patient Reference Frame (9732353) or Percutaneous Pin Adapter (9734752). Customers are instructed to: 1.Immediately locate and quarantine all unused impacted product(s). See Attachment A for affected lot numbers and product identification. 2.Return unused impacted product(s) to Medtronic following the instructions in the enclosed Customer Confirmation Form 3.Complete the Customer Confirmation Form enclosed with this letter, acknowledging that you have received this information, and email the completed form to Medtronic at neuro.quality@medtronic.com. If the affected devices have already been utilized and/or discarded, we still ask that you complete and return the Customer Confirmation Form detailing that information. 4.This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. Any questions regarding this communication, please contact your Medtronic Sales Representative or Technical Services at 1-888-826-5603.
Initiated
2024-12-04
Root cause
Under Investigation by firm
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, China, Costa Rica, Czech Republic, Denmark, Finland, Germany, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Mexico, Morocco, Namibia, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Reunion, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, United Arab Emirates, United Kingdom, Vietnam.
Recall number
Z-1022-2025
Recall status
Open, Classified
CDRH record ID
211670
FDA classified
2025-01-27
Recall event number
95970

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗