Biosense is recalling the 20-Pole Eco Cable because it was incorrectly calibrated. When a catheter is connected to an affected to an affected 20-Pole Eco Cable, the catheter will not be recognized by the impacted 54 cable serial numbers within the field.
Source facts
Action
Biosense Webster sent an Urgent Field Notice Medical Device Letter dated March 5, 2013 to all affected customers. The letter identified the affected product, problem and the actions to be taken. Customers with questions are instructed to contact Biosense Webster sales representatives or call (866) 473-7823, Mon-Friday from 7amto 8pm EST.
Initiated
2013-03-05
Root cause
Employee error
Terminated
2013-10-17
Distribution
Worldwide Distribution - USA (Nationwide) including the states of: DE, NL, CH, FR, GB, DK.
Recall number
Z-1023-2013
Recall status
Terminated
CDRH record ID
116525
FDA classified
2013-03-28
Recall event number
64579