Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.
Source facts
Action
Repligen will repair the connections in each potentially affected MAVERICK device. In
the meantime, all customers have been instructed to refrain from using the device.
Initiated
2025-12-10
Root cause
Radiation Control for Health and Safety Act
Distribution
U.S.
Recall number
Z-1024-2026
Recall status
Open, Classified
CDRH record ID
218071
FDA classified
2026-01-13
Recall event number
98260