There may be dissociation of the polyethylene from the baseplate because the inside diameter of the polyethylene articular bearing was manufactured to an oversized condition.
Source facts
Action
Hospitals were notified via letter dated June 10, 2005 requesting that they quarantine product on hand and send in a response card to the firm.
Initiated
2005-06-07
Root cause
Other
Terminated
2006-05-02
Distribution
Nationwide and international.
Recall number
Z-1027-05
Recall status
Terminated
CDRH record ID
39545
FDA classified
2005-07-21
Recall event number
32282