Open original FDA record ↗Recall record
- FDA identifier
- Z-1027-2011
- Company in FDA record
- Impac Medical Systems Inc
- FDA classification date
- Aug 21, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Potential for improper treatment. A software error may result in potential major underdosing to targeted areas from jaw blockage of the MLC aperture (jaw intrusion), and dosing to non-targeted areas from interstitial leaf leakage (jaw extrusion).
Source facts
- Action
- Impac Medical Systems Inc sent an Important Safety Notice letter to all affected consginees. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to contact Impact Software support as soon as possible if they have the affected configurations identified in the letter. Customers were asked to complete the form included with the letter and and fax to: European customers: +44 1293 654401 American & rest of world customers 702-992-5002 or e-mail to support@impac.com Customers were asked to distribute the Safety Notice to any and all users of IMPAC software at their organization who are potentially affected by this issue. For questions e-mail support@impac.com or call European customers: 44 1293 654320 American & rest of world customers: 800-488-4672.
- Initiated
- 2009-07-17
- Root cause
- Software design
- Terminated
- 2011-11-25
- Distribution
- Worldwide Distribution - USA including AL, CA, CT, FL, GA, IA, IL, KY, MI, MN, MS, ND, NJ, OH, OR, PA,TN, TX, VA, WA, WI, and WV and the country of Canada.
- Recall number
- Z-1027-2011
- Recall status
- Terminated
- CDRH record ID
- 84252
- FDA classified
- 2009-08-21
- Recall event number
- 52903