In case of a system crash, images may not be written to the hard disk and this may result in inconsistencies in the database. In case of a system crash (e.g. blue screen, power outage) images may not be written from cache to the hard disk and might get lost.
Source facts
Action
Siemens sent a Safety Advisory Notice dated January 13, 2015, to all affected customers. The Safety Notice identified the product, the problem, and the action to be taken. Siemens recommended checking patient data in the event of a system crash and advised customers that a software update is planned for mid-2015 that will resolve the issue.
For questions regarding this recall call 610-219-6300.
Initiated
2015-01-13
Root cause
Software Manufacturing/Software Deployment
Terminated
2017-02-16
Distribution
Nationwide Distribution
Recall number
Z-1027-2015
Recall status
Terminated
CDRH record ID
133015
FDA classified
2015-01-28
Recall event number
70318