During re-work of disposable instruments to relabel for 4-year shelf-life of disposable instruments, parts got mixed into other lots.
Source facts
Action
An email was sent out to the consignees on 03/31/2022. All customers were instructed to segregate any affected product and contact Restor3d. If all product has already been returned to restor3d, the consignee is to indicate that on the enclosed response form.
On 04/19/2022, the firm extended the recall to two additional lots. The recall notification was sent to the consignees via email with a request for all affected product to be returned.
Initiated
2022-03-30
Root cause
Process change control
Terminated
2023-08-28
Distribution
US Nationwide distribution in the states of KY, NC, NJ, TX, IN, VA and AL.
Recall number
Z-1027-2022
Recall status
Terminated
CDRH record ID
193282
FDA classified
2022-04-29
Recall event number
89983