Open original FDA record ↗Recall record
- FDA identifier
- Z-1027-2026
- Company in FDA record
- Remote Diagnostic Technologies Ltd.
- FDA classification date
- Jan 8, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Patient monitor measures Intracranial Pressure (ICP) and Bladder Pressure (BDR) but ICP and BDR measurements have not been tested or validated for accuracy. If healthcare providers base clinical judgment on unvalidated measurements, this could potentially lead to incorrect clinical decisions.
Source facts
- Action
- Correction notices were mailed on 11/26/2025 to customers were informed of the following: U.S.: Do not use the Intracranial Pressure or Bladder Pressure measurements. Firm intends to update labeling. OUS: Continue to use the Intracranial Pressure and Bladder Pressure measurements. Firm intends to conduct verification and validation testing. Firm recommends routine cross-verification of most ICP/BDR readings serves to reduce potential risk inherent with ICP/BDR use. Share the correction notice with all relevant personnel within your organization and with any organization where affected devices have been transferred. Distributors forward the notice to your customers. Store a copy of this notice with the User/Operator Manual. Complete and return the response form via email to ECR.Recall.Response@Philips.com For further information about this issue contact the firm at (800) 722-9377 or ECR.Recall.Response@Philips.com
- Initiated
- 2025-11-26
- Root cause
- Device Design
- Distribution
- Worldwide - US Nationwide distribution including in the states of AZ, VA, FL, CA, TX, NM, IA, NJ, TN, OH, KS, IN, NC, IL, WY, WI, NV, NY, NE, PA, AL, MA, GA, MO, CO, OK, KY, OR, MS, MI, HI, WA, UT, WV, ID, ND, NH, MN, MT, ME, MD, DE and the countries of AU, GB, IT, SA, DE, AE, BO, FR, ES, BE, IL, SG, CA, IE, TH, NL, MY, AT, DK, RO, CH, IS, PT, NO, HK, PL, KY, OM, CZ, SI, SL, GI, ZA, AF, SE, FI.
- Recall number
- Z-1027-2026
- Recall status
- Open, Classified
- CDRH record ID
- 216971
- FDA classified
- 2026-01-08
- Recall event number
- 97889