The caster/wheel may come off of the bassinet, causing the bassinet to tip.
Source facts
Action
Customer notification letters dated June 20, 2003 were sent to each customer. Customers are asked to verify the security of the casters to the base and to visually check for loose or missing caster screws, to remove any bassinets from service until repaired, and that Hill-Rom will visit each customer to replace the caster screws on all units.
Initiated
2003-06-20
Root cause
Other
Terminated
2003-12-19
Distribution
United States, Brazil, Canada, China, Italy, Japan, Kuwait Netherlands, Qatar, United Kingdom
Recall number
Z-1028-03
Recall status
Terminated
CDRH record ID
27937
FDA classified
2003-07-18
Recall event number
26586