Skip to content

Public FDA record

Z-1028-2017: Fresenius 2008K2 Series : Hemodialysis System

Recall record Z-1028-2017 for Fresenius Medical Care Renal Therapies Group, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1028-2017
Company in FDA record
Fresenius Medical Care Renal Therapies Group, LLC
FDA record creation date
Jan 13, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine

Source facts

Action
Fresenius issued Customer Notification via Certified Mail on 12/21/2016. Users provided with problem and recommendation, when adjusting the UF Rate, UF Goal, or Tx Time, is to use the numeric data entry keys instead of the up or down arrow keys. Change the values by using data entry keys, then save by pressing CONFIRM. Fresenius will make arrangements to install software on the hemodialysis machine. Confirmation of the noticed by the consignee will be required via a Reply Form.
Initiated
2016-12-21
Root cause
Device Design
Terminated
2023-03-17
Distribution
Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam
Recall number
Z-1028-2017
Recall status
Terminated
CDRH record ID
152020
FDA record created
2017-01-13
Recall event number
76037

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗