Skip to content

Public FDA record

Z-1028-2026: BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-0

Recall record Z-1028-2026 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1028-2026
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Jan 8, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due a software issue that may result in equipment not receiving timely updates of patient, medication, or related information from enterprise systems.

Source facts

Action
On 12/03/2025, the firm sent U.S. Postal and email an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them that BD received customer complaints of BD Pyxis" equipment not receiving timely updates of patient, medication, or related information from enterprise systems. This resulted in new patient and/or order information to be delayed from displaying on connected Pyxis equipment, and a delay in any changes to current patient orders being applied to the patient profile within Pyxis. Customers are instructed to: 1.Refer to the BD Pyxis" ES User Guides warning: Instruct end users to utilize the BD Pyxis" ES System patient/order information in conjunction with the medical record when making clinical decisions. Making clinical decisions solely from BD Pyxis" ES System information may result in errors in care resulting in serious injury or death. 2.BD Pyxis" MedStation" devices can be utilized in both profile mode (specific medications orders are associated with specific patients) and non-profiled mode. 3.Facilities should establish comprehensive policies and procedures for situations when an automated dispensing system is not receiving updated information and medications may not be loaded in device inventory. 4.Device keys can be used to manually access unavailable drawers. Keys must be readily available and retrievable to constitute risk mitigation. 5.Devices can also be placed into critical override mode, where all contents of the device will be available to users. (For additional detailed instruction - refer to Urgent Medical Device Correction letter. BD recommends that customers who are preparing to implement or upgrade to ES 1.11 continue as planned in order to receive the bug fixes it contains and prepare to receive the ES 1.11.1 patch when it is available. For Questions: BD Remediation Support: Email: US-Pyxis-Field-Action-Remediation@bd.com BD Technical Support: Website: bd.com/self-service, Phone: 1-800-727-6102, Hours: 24 hours, 7 days a week
Initiated
2025-12-03
Root cause
Under Investigation by firm
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK,AL,AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Bermuda, and Canada.
Recall number
Z-1028-2026
Recall status
Open, Classified
CDRH record ID
215904
FDA classified
2026-01-08
Recall event number
97546

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗