Z-1029-2011: SEQUENCER for Mosaiq interfaced to Electa SL75-5 machines. Sequencer Verify and Record System is to assi

Recall record Z-1029-2011 for Impac Medical Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1029-2011
Company in FDA record
Impac Medical Systems Inc
FDA classification date
Feb 4, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Computer interface is defected. Computer based verification system for recording patient treatment on a radiotherapy treatment machine fails to record treated field.

Source facts

Action
A User Notice was sent by the firm to its consignees. It identified the affected product and the reason for the recall. It also discussed the clinical impact and interim mitigations. The firm states in the letter that a fix for the problem and been developed and is available. Customers are to contact IMPAC Support for the patch to the SL75-5 interface. If customers have questions, concerns, or requests, they can email support@impac.com or one of the numbers listed in the letter for the applicable country. The letter should be distributed to all users affected by the issue at the customer's organization.
Initiated
2010-03-10
Root cause
Software design
Terminated
2011-04-06
Distribution
Worldwide Distribution -- US, Canada, Germany, and France.
Recall number
Z-1029-2011
Recall status
Terminated
CDRH record ID
90908
FDA classified
2011-02-04
Recall event number
55087

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗