Reason stated in the source
System software v1.5.1 has a system software issue related to the Daylight Savings Time (DST) change that can cause a system error, requiring the user to restart the system and potentially cause a delay in treatment
Source facts
- Action
- Siemens notified US Consignees contacted via Email/Field Engineers/Sales Reps starting March 11, 2022. Letter Urgent Medical Device Correction states reason for recall, health risk and action to take: The following instructions are to be followed prior to the first use after Daylight Savings Time 2:00 am US Eastern Standard Time, Sunday, March 13, 2022, regardless of local time and observance of Daylight Savings Time. This will synchronize the system clock. 1. Power on the System 2. Wait for the Home screen to appear The following instructions are to be followed prior to the first use after Daylight Savings Time 2:00 am US Eastern Standard Time, Sunday, March 13, 2022, regardless of local time and observance of Daylight Savings Time. This will synchronize the system clock. 1. Power on the System 2. Wait for the Home screen to appear. Once this notification is received, please complete, sign, and return the attached Business Response Form (page 3 of this letter) to OneMD-Field-Actions@its.jnj.com within 3 business days of receipt of this notification. Please include in the email subject: FA 2090694: 2022 VELYS" Station SW DST Issue. " Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). " Contact your VELYS" Technical Support Team through email at velystechsupport@its.jnj.com or via phone at 1-888-228-3597 for additional support to correct this issue.
- Initiated
- 2022-03-11
- Root cause
- Software design
- Terminated
- 2024-04-18
- Distribution
- Worldwide distribution - US Nationwide and the countries of Israel, Australia & New Zealand.
- Recall number
- Z-1029-2022
- Recall status
- Terminated
- CDRH record ID
- 192989
- FDA record created
- 2022-05-02
- Recall event number
- 89959