Possible error in the table longitudinal position.
Source facts
Action
Elekta sent an updated IMPORTANT NOTICE letter dated December 24, 2008, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer.
Elekta recommended that all users, and Eleckta field service engineers do regular QA checks to compare the movement of the treatment table in a given axis with the digital readout.
Elekta Service representatives would attach the screws with Loctite and calibrate the axis on these sites.
For questions regarding this recall call 770-670-2548.
Initiated
2008-03-10
Root cause
Other
Terminated
2012-08-09
Distribution
Nationwide Distribution and Puerto Rico
Recall number
Z-1030-2011
Recall status
Terminated
CDRH record ID
96671
FDA classified
2011-02-03
Recall event number
57570