Skip to content

Public FDA record

Z-1031-04: Computed Tomography, Models TSX-101A and TSX-021A.

Recall record Z-1031-04 for Toshiba American Med Sys Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1031-04
Company in FDA record
Toshiba American Med Sys Inc
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Erroneous patient data is entered. Previoius patient's demographics are recorded.

Source facts

Action
Firm issued Field Modification Instruction which describes the action to be taken by the firm''s own service technicians. Recall letters were sent to each site informing of the issue and provide caution regarding the problem prior to it''s getting fixed.
Initiated
2004-03-17
Root cause
Software design
Terminated
2011-11-01
Distribution
Nationwide.
Recall number
Z-1031-04
Recall status
Terminated
CDRH record ID
33470
FDA classified
2004-07-20
Recall event number
29284

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗