Open original FDA record ↗Recall record
- FDA identifier
- Z-1031-2009
- Company in FDA record
- Zimmer Inc.
- FDA classification date
- Mar 4, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Some of the torque wrenches apply less than the minimum torque to the screw locking cap when the torque indicators are aligned as described in the surgical technique.
Source facts
- Action
- The firm's sales force was notified by an Urgent: Device Recall letter dated 10/7/08, which instructed them to stop using the instrument and quarantine it immediately; physically count all affected product in their territory and record data on the Inventory Return Certification Form; fax a copy of the completed form to Zimmer Spine; and return the affected devices with the completed Inventory Return Certification Form. User accounts/implanting physicians were informed by an Urgent: Device Recall letter dated 12/10/08 and instructed to monitor their patients.
- Initiated
- 2008-11-07
- Root cause
- Process control
- Terminated
- 2009-12-22
- Distribution
- Worldwide Distribution --- USA including states of California, Minnesota, Ohio, Pennsylvania, and Tennessee, and countries of Denmark, Germany, Iran, Italy, Russia, Singapore, Spain, Switzerland, and United Kingdom.
- Recall number
- Z-1031-2009
- Recall status
- Terminated
- CDRH record ID
- 74904
- FDA classified
- 2009-03-04
- Recall event number
- 50176