the anti-toggle ring of the hex driver may separate from the driver shaft. If this occurs during surgery, the anti-toggle ring can be pressed into the pedicle screw head and cause separation of the screw from the housing upon tightening. if the anti-toggle ring remains in the pedicle crew head and can interfere with final tightening of the pedicle screw construct. this can lead to failure of the set crew and relies of the rod from the construct.
Source facts
Action
The anti-toggle ring of the HEX Driver, 1/4", square QC, 40 can separate form the driver shaft.
Initiated
2018-07-05
Root cause
Nonconforming Material/Component
Terminated
2021-04-28
Distribution
distributor and user level
Recall number
Z-1031-2019
Recall status
Terminated
CDRH record ID
171124
FDA classified
2019-03-20
Recall event number
82201