Skip to content

Public FDA record

Z-1032-2009: InteliFUSE, Drill Bit, 1.50 mm, J-Latch, QTY: 1, REF: DB.150.JL, non-sterile, 1515 Poydras st., Ste 1490

Recall record Z-1032-2009 for Intelifuse Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1032-2009
Company in FDA record
Intelifuse Inc
FDA classification date
Mar 5, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Manufacturing specification for drill bit was incorrect which, when used in conjunction with the drill guide, could cause the production of metal shavings.

Source facts

Action
The firm initiated their recall on July 31, 2008 via telephone and Urgent Recall Notice to their current distributor, Integra. The letter asked them to complete, sign, and return the enclosed Recall Acknowledgment and Return Form. The letter also stated that InteliFUSE would travel to their designated Distribution Center(s) to ensure testing and replacement of all defective units once the form has been received. The firm sent follow-up Urgent Recall Notice letters on August 14, 2008 directing the distributor to recall from the hospitals to whom they distributed the product and to two hospitals which were shipped the products directly. The letter asked that use of the affected products be immediately ceased and that the any stock on hand be quarantined. The Recall Acknowledgment and Return Form should be completed, signed, and returned and the products should be returned to InteliFUSE for immediate replacement.
Initiated
2008-07-30
Root cause
Process control
Terminated
2011-06-22
Distribution
Nationwide Distribution - NY and NJ.
Recall number
Z-1032-2009
Recall status
Terminated
CDRH record ID
73220
FDA classified
2009-03-05
Recall event number
49372

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗