Open original FDA record ↗Recall record
- FDA identifier
- Z-1032-2016
- Company in FDA record
- Novadaq Technologies Inc.
- FDA record creation date
- Mar 2, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It has been determined that the PINPOINT Operator's Manual does not contain validated parameters for gravity-cycle autoclave sterilization of the PINPOINT laparoscopes and light guide cable in the reprocessing tray.
Source facts
- Action
- Novadaq sent an Urgent Device Correction letter dated February 2, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Recipients of this notice have received shipments that include the inaccurate Operator's Manual. Please follow these instructions: 1. Discard any old PINPOINT Operator's Manuals Rev J through M. 2. Accept and review the new PINPOINT Operator's Manual (Rev N). 3. Complete and sign the enclosed "Acknowledgement of Receipt" form, scan it and email to PI NPOINT@novadaq.com. Any questions or concerns may also be sent to this address. For further questions, please call 1-(844) 668-2327, ( Select Option # 1)
- Initiated
- 2016-01-29
- Root cause
- Process design
- Terminated
- 2016-11-03
- Distribution
- Worldwide Distribution - Nationwide Distribution including Puerto Rico and to the countries of : Bangladesh, Canada, Hong Kong, India, and Israel.
- Recall number
- Z-1032-2016
- Recall status
- Terminated
- CDRH record ID
- 143596
- FDA record created
- 2016-03-02
- Recall event number
- 73225