Skip to content

Public FDA record

Z-1033-2015: DrugCheck Ketamine Dip Test 31317-3P PRO-EU-SWE, 31226-3-CA-GML, 30842 TR, 30108C, 30108C CAN, 30108C EU,

Recall record Z-1033-2015 for Express Diagnostics Int'l., Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1033-2015
Company in FDA record
Express Diagnostics Int'l., Inc.
FDA classification date
Jan 30, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

This recall has been initiated due to potential false positives results when using the test to screen for Ketamine. Use of this product may give incorrect screening results.

Source facts

Action
Consignees were sent an Express Diagnostic Int. Inc., "Urgent" Letter dated 11//21/2014. The letter described the problem and the product involved in the recall. It advised consignees to quarantine the product. If the product has been further distributed, consignees were advised to notify their customers at once, Consignees were requested to complete and return the "Acknowledgement Form" either via fax to 1-507-526-2252 or e-mail Quality@drugcheck.com. For questions consignees can contact Jackie Gale, Director of Quality Systems and Regulatory Affairs at 1-507-526-3951.
Initiated
2014-11-21
Root cause
Nonconforming Material/Component
Terminated
2019-05-14
Distribution
Worldwide Distribution -- US, including the states of CA, NY, MN, and SC; SWEDEN, CANADA, UNITED KINGDOM, and GUATEMALA.
Recall number
Z-1033-2015
Recall status
Terminated
CDRH record ID
132104
FDA classified
2015-01-30
Recall event number
70008

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗