Potential for sterile blades, within non-sterile SPT kits containing the blades, to puncture the outer foil layer causing a breach in the sterile packaging.
Source facts
Action
Medline Industries, LP notified consignees via email and first-class mail on 11/21/2022. Consignees were instructed to over-label the sterile kits with a warning label indicating Warning To Avoid Injury, Please Handle Blades with Care . For the Non-Sterile kits, they are to over label the affected kits indicating to remove the blades and replace from a different sterile supply. No product is to be returned. Customers were also instructed to complete and return the response form.
Initiated
2022-11-17
Root cause
Device Design
Distribution
US Nationwide distribution in the states of AR, MD, WA, WI.
Recall number
Z-1033-2023
Recall status
Open, Classified
CDRH record ID
197711
FDA classified
2023-01-27
Recall event number
91387