Some of the high voltage capacitors in the recalled external defibrillators have had extended charge times resulting in self-test failures following which the defibrillator cannot be used/cannot be turn on.
Source facts
Action
A recall letter, dated May 24, 2004, stated that a replacement AED is enclosed with the letter, and instructed consignees to replace their AED with the new one immediately and return the old AED.
Initiated
2004-05-24
Root cause
Other
Terminated
2005-11-18
Distribution
The recalled AED units were distributed nationwide in the United States and to foreign accounts in Europe,
Asia, Latin America, the Middle East, and Africa.
Recall number
Z-1034-04
Recall status
Terminated
CDRH record ID
33195
FDA classified
2004-07-20
Recall event number
29024
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.