Open original FDA record ↗Recall record
- FDA identifier
- Z-1034-2015
- Company in FDA record
- Animas Corporation
- FDA classification date
- Jan 30, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Calibration factors in the pump overwritten during a programming step. The force sensor could send a lower signal value to the pump processor, with loss of prime warnings, occlusion alarms and the pump unable to detect a cartridge during the prime sequence. Field action initiated 8/29/2011.
Source facts
- Action
- The field action was initiated on 8/29/2011 and comprised communication to patients and distributors using email, letters and verbal communication. In addition, notifications were sent to those Health Care Professionals who had patients that were affected by this field action. Health Authorities were notified in those countries where the pump had been distributed (France, Germany, Sweden and United Kingdom) or was in the hands of patients. Replacement product was provided to distributors and end users who were identified as having affected product.
- Initiated
- 2011-09-06
- Root cause
- Software design (manufacturing process)
- Terminated
- 2015-08-10
- Distribution
- No US distribution, Distributors are located in France, Germany, Sweden and United Kingdom.
- Recall number
- Z-1034-2015
- Recall status
- Terminated
- CDRH record ID
- 130189
- FDA classified
- 2015-01-30
- Recall event number
- 69299