Public FDA record

Z-1034-2026: Brand Name: HFD100 Product Name: Head Fixation Device HFD100 Model/Catalog Number: 119695-000, 119629-000

Recall record Z-1034-2026 for IMRIS Imaging Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1034-2026
Company in FDA record
IMRIS Imaging Inc
FDA classification date
Jan 8, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Torque screw may crack or separate. If used, an affected screw could lead to inadequate head stabilization, compromised surgical precision, procedural delays, requirement for alternate head fixation methods, and risk of patient injury.

Source facts

Action
FIrm notified consignees via an Urgent Medical Device Recall letter on December 8, 2025. Service will contact customers to replace the torque screw components. Until the torque screws are replaced, customers should follow these mitigations: " Ensure the patient s head is supported during pin penetration: serious patient injury may result from improper head and neck support during pin penetration of the skull for head placement. " If a crack develops during advancement or retraction, please discontinue use and release the skull clamp halves through activating the base locking knob. Please notify IMRIS. " Prior to each device use, closely inspect the torque screw component for any signs of wear, cracking, or damage along the exterior threads. " Handle the torque screw with care during assembly, disassembly, cleaning and storage between uses; if the torque screw is inadvertently dropped or impacted by another object, immediately discontinue use and notify IMRIS. " Ensure the instructions in the Operator Manual for proper HFD100 Head Fixation Device assembly and use are followed to produce rigid fixation. " Please distribute this notice to any personnel responsible for handling, cleaning, or storing HFD100 Head Fixation Device components and report any undesired device operation or feedback.
Initiated
2025-12-08
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.
Recall number
Z-1034-2026
Recall status
Open, Classified
CDRH record ID
217683
FDA classified
2026-01-08
Recall event number
98132

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗