Skip to content

Public FDA record

Z-1035-2026: Medline Hudson RCI Medium Concentration Oxygen Masks: 1041CE MASK,OXYGEN,UND CHIN,MED CO OUS ONLY; HUD10

Recall record Z-1035-2026 for Medline Industries, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1035-2026
Company in FDA record
Medline Industries, LP
FDA classification date
Jan 8, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Firm received multiple complaints regarding tubing disconnecting from oxygen mask. If issue occurs during use, may lead to delay in patient care and patient may experience shortness of breath leading to hypoxia, which may require medical intervention to prevent further respiratory deterioration.

Source facts

Action
On November 21, 2025, the firm notified customers via Urgent Medical Device Recall letter. Customers were instructed to destroy affected product on hand. No product is being returned to Medline. For kits, Medline instructed customers to request stickers to over-label the affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply.
Initiated
2025-11-21
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Bahamas, Canada, China, Dominican Republic, Guam, South Korea, Mexico, Panama, and Singapore.
Recall number
Z-1035-2026
Recall status
Open, Classified
CDRH record ID
217666
FDA classified
2026-01-08
Recall event number
98125

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗