Open original FDA record ↗Recall record
- FDA identifier
- Z-1035-2026
- Company in FDA record
- Medline Industries, LP
- FDA classification date
- Jan 8, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Firm received multiple complaints regarding tubing disconnecting from oxygen mask. If issue occurs during use, may lead to delay in patient care and patient may experience shortness of breath leading to hypoxia, which may require medical intervention to prevent further respiratory deterioration.
Source facts
- Action
- On November 21, 2025, the firm notified customers via Urgent Medical Device Recall letter. Customers were instructed to destroy affected product on hand. No product is being returned to Medline. For kits, Medline instructed customers to request stickers to over-label the affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply.
- Initiated
- 2025-11-21
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Bahamas, Canada, China, Dominican Republic, Guam, South Korea, Mexico, Panama, and Singapore.
- Recall number
- Z-1035-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217666
- FDA classified
- 2026-01-08
- Recall event number
- 98125