One of three copper spectral filters could remain within the x-ray while the system displays no spectral filter. The displayed Air Kerma rate and thus cumulative Air Kerma are overestimated. The Air Kerma rate may be over estimated by up to 60%.
Source facts
Action
Affected customers were notified by means of a Important Electric Product Radiation Warning letter. GE field personnel will be replacing the collimators.
Initiated
2008-11-04
Root cause
Radiation Control for Health and Safety Act
Terminated
2012-05-16
Distribution
Nationwide Distribution -- AL, AZ, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV and Puerto Rico.
Recall number
Z-1036-2009
Recall status
Terminated
CDRH record ID
84852
FDA classified
2009-09-24
Recall event number
53176