Open original FDA record ↗Recall record
- FDA identifier
- Z-1036-2010
- Company in FDA record
- Pointe Scientific, Inc.
- FDA classification date
- Mar 9, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
An additional 10% picric acid was added to the R2 component during production.
Source facts
- Action
- Consignees were notified by An Urgent: Medical Device Recall letter, dated 12/1/09, that identified the affected product. Customers were asked to immediately examine their inventory for the affected product and to notify any customers that the product may have been further distributed to by giving them a copy of the recall notice. Customers are to dispose of the affected product immediately. Questions should be directed to the Technical Support Department at (800)757-5313 or (734)487-8300. There is also a Recall Response form attached with this notice.
- Initiated
- 2009-12-01
- Root cause
- Process change control
- Terminated
- 2010-10-15
- Distribution
- Worldwide Distribution -- USA, Bolivia, Greece, Guyana, India, Indonesia, Israel, Kenya, Korea, Malaysia, Mexico, Pakistan, Philippines, Russia, Serbia, Sri Lanka, Tanzania, Trinidad and Ukraine.
- Recall number
- Z-1036-2010
- Recall status
- Terminated
- CDRH record ID
- 87473
- FDA classified
- 2010-03-09
- Recall event number
- 54094