The 8mm and 10mm Rotating Cutters may fracture during use, specifically during disc space preparation. If an 8mm or 10mm Rotating Cutter does Fracture during use, it may result in a surgical delay estimated to be 30 minutes or greater due to preparing the disc space using other instruments or methods and/or retrieving the instrument fragment.
Source facts
Action
Orthofix mailed recall letters to the distributors on 02/06/2012. Orthofix does not sell or consign these instruments to hospitals.
Initiated
2012-02-06
Root cause
Device Design
Terminated
2012-06-11
Distribution
Nationwide and Puerto Rico
Recall number
Z-1037-2012
Recall status
Terminated
CDRH record ID
107325
FDA classified
2012-02-15
Recall event number
61117