Skip to content

Public FDA record

Z-1037-2025: HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx15mm UPN: M00553570

Recall record Z-1037-2025 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1037-2025
Company in FDA record
Boston Scientific Corporation
FDA classification date
Jan 29, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.

Source facts

Action
Boston Scientific issued Urgent Medical Device Removal - Immediate Action Required Letter via Fedex Priority Overnight on December 20, 2024 to US consignees. Letter states reason for recall, health risk and action to take: This removal affects only the products listed in Attachment 1. Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. Share this notice with any health care professional within your organization and with any organization where the affected products may have been transferred. " If you are a distributor, this notification must be forwarded to your customers to ensure notification of this product removal is carried out to the end-user level. If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them. Contact your local Boston Scientific representative with any questions.
Initiated
2024-12-20
Root cause
Process control
Distribution
Nationwide including Puerto Rico Foreign: To be provided
Recall number
Z-1037-2025
Recall status
Open, Classified
CDRH record ID
211878
FDA classified
2025-01-29
Recall event number
96049

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.