- FDA identifier
- Z-1038-2010
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA classification date
- Mar 10, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Results can elevate after sensor replacement. Sodium urine results can be falsely elevated 15 mmol/L within 24 hours after installation of a new sensor.
Source facts
- Action
- The recalling firm issued an Urgent Field Safety Notice dated October 2009 to all affected customers. Customers were given instruction which minimize urine specimen use and report urine specimen results when QC is acceptable. Sensors are to be replaced when urine QC is out. Replacement sensors are available by contacting the firm at 800-241-0420. Please contact the Siemens Technical Solutions Center at 800-441-9250 for further assistance.
- Initiated
- 2009-10-22
- Root cause
- Other
- Terminated
- 2010-12-16
- Distribution
- The product was shipped to medical facilities nationwide. The product was also shipped to Australia, Canada, Belgium, France, Germany, Italy, Japan, Malaysia, New Zealand, Netherlands, Portugal, Spain, South Korea, and Switzerland.
- Recall number
- Z-1038-2010
- Recall status
- Terminated
- CDRH record ID
- 86322
- FDA classified
- 2010-03-10
- Recall event number
- 53686