Open original FDA record ↗Recall record
- FDA identifier
- Z-1038-2012
- Company in FDA record
- European Custom Manufacturing B.V.
- FDA classification date
- Feb 15, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The crimping operation during product production was not performed properly. As such, the pacing wire may separate from the connection pin, which causes non-functionality.
Source facts
- Action
- ECM sent an "ECMbv RECALL OF MCE TEMPORARY PACING WIRES" letter dated May 2, 2011 to all affected customers. Customers were also informed of a potential problem via e-mail and/or phone between April 18 and 21, 2011. The letter describes the product, problem, and actions to be taken by the customers. Customers were instructed to place the affected product on hold and to return the product to the firm. For questions on this matter, contact ECM or MCE Customer Service at 1-877-886-1985.
- Initiated
- 2011-04-18
- Root cause
- Process design
- Terminated
- 2012-02-15
- Distribution
- Nationwide Distribution-including the states of AL, AR, CA, FL, GA, KY, NC, NV, NY, TX, VA, and WI.
- Recall number
- Z-1038-2012
- Recall status
- Terminated
- CDRH record ID
- 106845
- FDA classified
- 2012-02-15
- Recall event number
- 60942