Reason stated in the source
The interface fails to send abnormal flags for Reference Lab test results.
Source facts
- Action
- The firm sent out 2 CORRECTION COMMUNICATIONS informing consignees of functionality in release of 4.0.7 HIS Add-on interfaces which SCC has designated to be a potential patient Risk-to-Health issue. Further the corrective action indicated HIS releases 4.0.7.2 and higher will be corrected. Hot Fixes have been developed for HIS Add-on patches 4.0.7.0 and 4.0.7.1. Lastly the communication requests that the user "please acknowledge receipt of this task and grant permission to load the mandatory hot fix(es). Thank you."
- Initiated
- 2014-05-20
- Root cause
- Device Design
- Terminated
- 2016-05-16
- Distribution
- Distributed US (nationwide) in the states of OH, GA, NJ, WI, FL, NH, IL, NY, IN, CA, AL, VA, MO, MI, KY, MA, PA, CO, ME, TN, WA, MN, SD, LA, NC, and NV and the country of Canada.
- Recall number
- Z-1038-2015
- Recall status
- Terminated
- CDRH record ID
- 132127
- FDA classified
- 2015-02-02
- Recall event number
- 70021