Skip to content

Public FDA record

Z-1038-2023: Turbett Surgical Container, TS1000: Indicated for enclosing other medical devices that are to be steriliz

Recall record Z-1038-2023 for Turbett Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1038-2023
Company in FDA record
Turbett Surgical, Inc.
FDA classification date
Jan 30, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Pod may have weld failures breaching the sterile barrier and result in the use of unsterile instruments leading to infection or other major injuries.

Source facts

Action
***UPDATE*** Turbett Surgical issued via E-mails Product Alert directly to the Hospital departments utilizing the device on 12-27-2022. Letter states reason for recall, health risk and action to take: Actions to be taken by the Customer/User: 1. Immediately discontinue the use of the specific devices exhibiting one or more failure(s) listed below. " Missing side supports " Broken or cracked side tab support " Backwall weld separation " Side weld separation " Crack in the backwall seam 2. Email (quality@turbettsurgical.com) with the following information. " Hospital name " Serial number(s) " Type of failure that your container exhibited 3. Turbett Surgical will contact you to organize pick up of the damaged unit and shipment of a replacement unit. If your container does not exhibit any of these failures, please continue to monitor and inspect the condition of the container and follow the Instructions For Use (IFU) as you continue utilizing your container. 4. If your container does not exhibit any of these failures, please email quality@turbettsurgical.com stating that your container(s) do not exhibit any of these failures. Please continue to monitor and inspect your devices prior to assembling for sterilization AS WELL AS after sterilization and cooling to ensure expected outcomes. Reminder to follow the Instructions For Use (IFU) as you continue utilizing your TS-1200 container. If you have any questions about the content of this letter, please contact Mark Petras at quality@turbettsurgical.com. ________________________________________________________________________ Turbett Surgical issued Emails to the Hospital department on 12/7/2022. Sales representatives were notified on 12/5/22 to switch out the other container. Letter states reason for recall, health risk and action to take: 1. Immediately verify that the specific serial numbered devices listed below, continues to operate without creating wet loads or the sterilization process is not compromised.
Initiated
2022-12-05
Root cause
Device Design
Distribution
US Nationwide distribution in the states of CA, IA, IN, NC, NH, NY, OH, LA, SD, TX, VA, WI.
Recall number
Z-1038-2023
Recall status
Open, Classified
CDRH record ID
197435
FDA classified
2023-01-30
Recall event number
91314

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗