Incorrect Sterilization Process --The Operating Manual for the product contains an incorrect statement regarding the sterilization of the Autolube foot control. The manual states that the product can withstand ETO sterilization, but there is no data to support this process.
Source facts
Action
Anspach issued an Urgent Device Correction letter on 11/27/06, to all manual holders advising them of the error and requested that they update the manual with the corrected instructions. A reply receipt confirming notification was included.
Initiated
2006-12-18
Root cause
Labeling False and Misleading
Terminated
2012-07-06
Distribution
Worldwide Distribution
Recall number
Z-1039-2008
Recall status
Terminated
CDRH record ID
50061
FDA classified
2008-02-13
Recall event number
37183