- FDA identifier
- Z-1039-2019
- Company in FDA record
- Shimadzu Medical Systems Usa Com
- FDA record creation date
- Mar 21, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
In these units, the brake is usually released by gripping the drive handle and applied by letting go of the handle. However, for systems within a specified manufacture period, we have identified that even when letting go of the handle, the systems brakes are not applied. In this event, if the Operator releases the handle when the device is moving, the unit will continue to coast.
Source facts
- Action
- Shimadzu Medical Systems sent an Urgent Medical Device Correction Notice letter dated January 7, 2019 to customers. The letter identified the affected product, problem and actions to be taken. The corrective action for systems identified with this issue is replacement of the systems drive handle. Customers will be contacted by a Shimadzu or Fuji local Service Representative to determine whether your system requires the above correction action. For questions contact the Shimadzu National Service Support group at (844) 487-2767.
- Initiated
- 2019-01-14
- Root cause
- Device Design
- Terminated
- 2023-04-13
- Distribution
- US Nationwide Distribution
- Recall number
- Z-1039-2019
- Recall status
- Terminated
- CDRH record ID
- 170357
- FDA record created
- 2019-03-21
- Recall event number
- 81940