Skip to content

Public FDA record

Z-1039-2023: AxiEM" Non-Invasive Patient Tracker

Recall record Z-1039-2023 for Medtronic Navigation, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1039-2023
Company in FDA record
Medtronic Navigation, Inc.
FDA classification date
Jan 30, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to increasing complaint trend for the Non-Invasive Patient Tracker system where complaints analysis indicates that users are unable to successfully verify their navigation instruments during image-guided surgery.

Source facts

Action
On 12/15/2022, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via mail courier informing customers is recalling two specific lots of AxiEM Non-Invasive Patient Trackers due to the potential for verification failure of navigation instruments during a Functional Endoscopic Sinus Surgery procedure. Customer are instructed: 1. Do NOT use any impacted product. Immediately locate and quarantine all unused impacted product(s). Refer to the affected lot numbers identified in Table 1 below. 2. Return the impacted product(s) to Medtronic following the instructions in the enclosed Customer Confirmation Form. Their Medtronic Sales Representative can assist in initiating the return of the product as necessary. 3. Complete the Customer Confirmation Form enclosed with this letter (even if they have no product to return), acknowledging that they have received this information, and email the completed form to Medtronic at neuro.quality@medtronic.com. 4. If the affected devices have already been utilized and/or discarded, the firm is still asking that customers complete and return the Customer Confirmation Form detailing that information. For questions, contact Medtronic Sales Representative or Medtronic Technical Services at 1-888-826-5603 or via email at rs.navtechsupport@medtronic.com.
Initiated
2022-12-15
Root cause
Process control
Distribution
Worldwide - U.S. distribution including in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV. The country of Japan.
Recall number
Z-1039-2023
Recall status
Open, Classified
CDRH record ID
197556
FDA classified
2023-01-30
Recall event number
91354

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.