Z-1039-2026: Medicrea, IB3D PL Instruments Set, Rx Only REF: SPS03174

Recall record Z-1039-2026 for Medicrea International. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1039-2026
Company in FDA record
Medicrea International
FDA classification date
Jan 9, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to complaints, orthopedic rotation tool handle may untread from the shaft resulting in the inability to rotate implant.

Source facts

Action
On October 29, 2025, Medtronic issued a Urgent: Medical Device Recall Notification via UPS. Medtronic ask consignees to take the following actions: 1. Immediately identify and quarantine the potentially affected product and instruments set identified in the table above. 2. Your local Medtronic Spine Representative will be contacting you to coordinate the return of potentially affected products and instruments sets. 3. Complete and return the Customer Confirmation Form, even if you do not have any affected product. Please complete all fields, sign the form, and email neuro.quality@medtronic.com. 4. This notice should be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. 5. If purchased from a distributor, contact your distributor directly to arrange for the return of the product to your Distributor
Initiated
2025-10-29
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of TN, CA, MN, NC, TX, FL, OH, RI, VA.
Recall number
Z-1039-2026
Recall status
Open, Classified
CDRH record ID
217404
FDA classified
2026-01-09
Recall event number
97996

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗