Open original FDA record ↗Recall record
- FDA identifier
- Z-1040-2015
- Company in FDA record
- Sedecal S.A.
- FDA classification date
- Feb 2, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Reports of unexpected movement.
Source facts
- Action
- URGENT - Field Safety Notice letters dated August 6, 2014 were sent to all consignees of record. The letters instructed consignees: 1) The unit is equipped with a safety switch in the drive handle. If the drive handle is released the unit will stop; 2) Do not try to correct unintended movement using the controls on the drive handle; 3) Instead, stop the system immediately by releasing the drive handle entirely; 4) Alternatively, the emergency button can be pressed; 5) Take special care when the system crosses metal strips or plates on the floor or enters an elevator, as these are situations where electrostatic discharge is likely to occur; and, 6) Should the customer feel uncertain regarding this action, he/she can contact Philips Healthcare.
- Initiated
- 2014-08-06
- Root cause
- Other
- Terminated
- 2016-06-22
- Distribution
- US: Nationwide (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NM, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI)
- Recall number
- Z-1040-2015
- Recall status
- Terminated
- CDRH record ID
- 132614
- FDA classified
- 2015-02-02
- Recall event number
- 70237