FUJIFILM Medical Systems U.S.A., Inc., Modality Systems became aware of the possibility that the braking mechanism might not be engaged in all cases when the driving handle has been released.
Source facts
Action
Fujifilm sent an Urgent Medical Device Correction and Removal letter dated January 16, 2019 to customers. The letter identified the affected product, problem and actions to be taken. Customers were advised to schedule a visit by our Service department to replace the handle assembly. Customers were provided with additional instructions to continue using the affected product safely until the correction was made. For questions call 201-675-8947.
Initiated
2019-01-15
Root cause
Device Design
Terminated
2020-10-08
Distribution
US Nationwide Distribution
Recall number
Z-1040-2019
Recall status
Terminated
CDRH record ID
171105
FDA record created
2019-03-21
Recall event number
82187