Open original FDA record ↗Recall record
- FDA identifier
- Z-1040-2023
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Jan 30, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
In rare cases, the system may only boot into backup mode after an abrupt shutdown and not reach full operating mode. This may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system.
Source facts
- Action
- On December 16, 2022, the firm initiated the recall to its affected customers via Urgent Medical Device Correction letters. Customers were instructed to do the following: Observe if the system suffers an abrupt shutdown. This can be caused e.g., by usage of the Emergency SHUTDOWN button, manually switching off via End session / Shutdown System command in the main menu, or by a long press on the power-on push button on the system console. In the event of an abrupt shutdown, restart immediately to verify system state and call service if necessary. The firm will provide a software patch to affected systems. Siemens's service organization will contact customers to arrange a date to perform the corrective action. Please feel free to contact the service organization for an earlier appointment at 1-800-888-7436.
- Initiated
- 2022-12-16
- Root cause
- Software design
- Distribution
- Worldwide - US Nationwide distribution.
- Recall number
- Z-1040-2023
- Recall status
- Open, Classified
- CDRH record ID
- 197697
- FDA classified
- 2023-01-30
- Recall event number
- 91418