Z-1040-2023: System Material # Artis zee floor 10094135 Artis zee ceiling 10094137 Artis zee multi-purpose 10094139

Recall record Z-1040-2023 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1040-2023
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Jan 30, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In rare cases, the system may only boot into backup mode after an abrupt shutdown and not reach full operating mode. This may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system.

Source facts

Action
On December 16, 2022, the firm initiated the recall to its affected customers via Urgent Medical Device Correction letters. Customers were instructed to do the following: Observe if the system suffers an abrupt shutdown. This can be caused e.g., by usage of the Emergency SHUTDOWN button, manually switching off via End session / Shutdown System command in the main menu, or by a long press on the power-on push button on the system console. In the event of an abrupt shutdown, restart immediately to verify system state and call service if necessary. The firm will provide a software patch to affected systems. Siemens's service organization will contact customers to arrange a date to perform the corrective action. Please feel free to contact the service organization for an earlier appointment at 1-800-888-7436.
Initiated
2022-12-16
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-1040-2023
Recall status
Open, Classified
CDRH record ID
197697
FDA classified
2023-01-30
Recall event number
91418

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗