Open original FDA record ↗Recall record
- FDA identifier
- Z-1041-2023
- Company in FDA record
- Carbon Medical Technologies, Inc.
- FDA classification date
- Jan 30, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The label incorrectly identifies the product as STAR1401, a 1x3mm Barbell shaped marker pre-loaded in a 14 gauge needle, while the packaging contains STAR1402, a 1x5mm Tribell shaped marker pre-loaded in a 14 gauge needle.
Source facts
- Action
- Carbon Medical Technologies sent an IMPORTANT: MEDICAL DEVICE RECALL notice by email to its sole consignee on 12/16/2022. The notice explained the issue and requested that the consignee perform the following actions: "Immediately examine your inventory and quarantine product subject to recall. In addition, if you may have further distributed this product, please identify you customers and notify them at once of this product recall." "Affected product should be returned to: Regulatory Affairs, Carbon Medical Technologies, Inc., 1290 Hammond Road, St. Paul, MN 55110-5959" "If you have any question, comments, or concerns please contact Customer Support at +1 888-207-2062. You may also email info@carbonmed.com."
- Initiated
- 2022-12-16
- Root cause
- Under Investigation by firm
- Terminated
- 2023-08-09
- Distribution
- US Nationwide distribution in the state of OHIO.
- Recall number
- Z-1041-2023
- Recall status
- Terminated
- CDRH record ID
- 197696
- FDA classified
- 2023-01-30
- Recall event number
- 91419