Open original FDA record ↗Recall record
- FDA identifier
- Z-1041-2026
- Company in FDA record
- Mobius Imaging, LLC
- FDA classification date
- Jan 9, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.
Source facts
- Action
- An URGENT MEDICAL DEVICE CORRECTION notice dated 12/4/25 instructs consignees to ensure the following mitigations are in place while the correction is deployed: the hand position warnings as indicated in the IFU, train operators, the AIRO is accessible through a physical key switch and password protected log in screen, and the Backing Up System Warning label is affixed on to the gantry of affected devices. Stryker will correct devices during an upcoming service visit. Consignees are to inform users of devices of this recall by sharing the recall notice, immediately check stock areas and operating rooms to identify affected devices, and inform all users of the potential harms and their risk mitigations. Consignees with any questions are to contact their local sales office or Stryker Representative.
- Initiated
- 2025-12-04
- Root cause
- Labeling Change Control
- Distribution
- US Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DC, FL, IL, KY, MA, MD, MI, MN, MO, NC, NY, OH, PA, RI, SC, TN, TX, UT, VA, WI.
- Recall number
- Z-1041-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217620
- FDA classified
- 2026-01-09
- Recall event number
- 98099